Probe
Type |
Part
Number
|
Measuring Range1
|
Accuracy
|
Resolution Steps
|
Range Steps
|
Metric (µm)
|
Imperial (mils/thou)
|
Metric (µm)
|
Imperial (mils/thou)
|
Metric (µm)
|
Imperial (mils/thou)
|
Metric (µm)
|
Imperial (mils/thou)
|
F1
Standard
|
|
T35511952
|
0
–1500
|
0
– 60
|
±1%
or ±1µm
|
±1% or ±0.04mil
|
0.1
0.05
1.0
|
0.005
0.02
0.05
|
0 - 200
200 -500
500 - 1500
|
0 - 8
8 - 20
20 - 60
|
F1
Right Angle
|
|
T35511953
|
F1
Telescopic
|
|
T35511959
|
F1
A (Automotive)
|
|
T35512400
|
F2
Standard
|
|
T35511954
|
0 - 5mm
|
0 - 200
|
±1% or ±5µm
|
±1% or ±0.2mil
|
2
5
|
0.1
0.2
|
0 - 0.5mm
0.5-5mm
|
0 - 20
20 - 200
|
F2
Right Angle
|
|
T35511955
|
F2
Telescopic
|
|
T35511960
|
F3
Standard
|
|
T35511956
|
0 - 13mm
|
0 - 500
|
±2% or ±30µm
|
±2% or ±1mils
|
5
10
|
0.2
0.5
|
0 - 1mm
1 - 13mm
|
0 - 40
40 - 500
|
F4
Standard
|
|
T35511950
|
0 - 250
|
0 - 10
|
±1% or ±1µm
|
±1% or ±0.04mil
|
0.1
|
0.005
|
0 - 250
|
0 - 10
|
F4
Right Angle
|
|
T35511951
|
F5
(Rebar)
|
|
T35511962
|
0 - 800
|
0 - 32
|
±1% or ±2µm
|
±1% or ±0.08mil
|
1
|
0.1
|
0 - 800
|
0 - 32
|
F6
Standard
|
|
T35511964
|
0 - 25mm
|
0 - 1000
|
±2% or ±100µm
|
±2% or ±4mils
|
10
50
|
0.5
2
|
0 - 5mm
5 - 25mm
|
0 - 200
200 - 1000
|
N1
Standard
|
|
T35511982
|
0 -1500
|
0 - 60
|
±1% or ±1µm
|
±1% or ±0.04mil
|
0.1
0.05
1.0
|
0.005
0.02
0.05
|
0 - 200
200 -500
500 - 1500
|
0 - 8
8 - 20
20 - 60
|
N1
Right Angle
|
|
T35511983
|
N2
Standard
|
|
T35511984
|
0 - 5mm
|
0 - 200
|
±1% or ±15µm
|
±1% or ±0.6mil
|
2
5
|
0.1
0.2
|
0 - 0.5mm
0.5-5mm
|
0 - 20
20 - 200
|
N4
(Anodisers)
|
|
T35511980
|
0 - 250
|
0 - 10
|
±1% or ±1µm
|
±1% or ±0.04mil
|
0.1
|
0.005
|
0 - 250
|
0 -10
|
1 all
measurements are displayed in either µm
or mils unless otherwise stated |